Loading Sample Plates in the Mass Spectrometer 3.4.1 Ejecting the Sample Holder In the Voyager Instrument Control Panel, select Eject from the Sample Plate menu. The Load/Eject dialog box is displayed. . Figure 3-4 Load/Eject Sample Plate Dialog Box The following also occurs: • A Load Status dialog box is displayed during the ejection sequence which displays hardware status. • High voltage is turned off. • The video monitor displays the sample plate moving. • The sample holder moves out of the main source chamber, out of the sample loading chamber, and is ejected from of the instrument. NOTE: Do not leave the sample holder in the Eject position for more than a few seconds. Minimize the time the instrument is exposed to the atmosphere to reduce the time needed to reach high vacuum when you insert a new plate. Voyager™ Biospectrometry™ Workstation User’s Guide 3-41 3