Impact of Changing Instrument Settings Parameters
Input Bandwidth
(not available with
Signatec
digitizers)
Decreasing the Input Bandwidth decreases the response time
of the detector, and can reduce higher frequency noise. A
lower setting can improve resolution and smooth out a
baseline, but may also decrease signal-to-noise ratio.
Because the Input Bandwidth is an electronic analog filter, it
may slightly shift peak centroids toward higher masses relative
to peaks recorded with Full Bandwidth.
Suggested settings
Use the settings listed below as a starting point for Input
Bandwidth. Adjust as needed to eliminate noise.
NOTE: To maintain signal-to-noise ratio, use a lower Input
Bandwidth setting.
Input Bandwidth
Mass Range
(Da)
Linear Mode Reflector Mode
0–500
Full
Full
500–6,000
200–250 MHz
Full
6,000–50,000
200–250 MHz
200–250 MHz
>50,000
20–25 MHz
20–25 MHz
PSD Mode
200–250 MHz
or 20–25 MHz
on PSD
segments
below 0.3
Mirror Ratio
NOTE: Settings below 200 in Linear mode may not allow
resolution of isotopes.