Download Bacharach Model 200 Specifications
Transcript
MODEL 300 COMBUSTION ANALYZER 3.1 PERFORMING AN EFFICIENCY TEST This procedure describes how to initially calibrate the Model 300 Combustion Analyzer and perform a combustion efficiency test. Each time the combustion analyzer is turned on, its microprocessor begins its initial start-up routine. This routine directs the operator to calibrate the analyzer and select the proper fuel. After performing an efficiency test, the analyzer can be recalibrated and/ or a different fuel can be selected for the next test. Refer to Sections 3.6 and 3.7. 1. Turn on the analyzer by either plugging it in, or by setting its battery pack switch to ON. WARNING! Use only UL or CSA approved grounded extension cords. Use of other cords may bring about a hazardous condition and voids the instrument’s warranty. Ungrounded extension cords may cause erratic display indications. 2. During initial power up, all front panel, and Optional auxiliary panel displays and LED indicators are checked by being turned on for approximately 5 seconds; after which, the left display switches to show CAL, the Optional clock displays the correct time, and the Optional auxiliary displays go dark. Also at this time, the pump is turned on and continues to run until the analyzer is calibrated. 3. Place probe into center of furnace’s primary air inlet (Section 3.2). If the furnace is drawing in ambient room air, you can sample room air for the purpose of determining the primary air temperature. NOTE: During calibration, you can measure inlet air temperature by selecting TEMP on the left display. Instruction 24-9223 3-3