25 Measure Commands
:MEASure:PTOP
Command
:MEASure:PTOP <source>, <pulse width percent>
The :MEASure:PTOP command measures the average of the data of a
positive pulse within the pulse window. The pulse window is a range of
data centered within the pulse width using the specified percentage of the
data as measured as the middle threshold level. For example, a 50%
window would not include in the average the first or last 25% of the pulse
width as measured at the middle threshold level. A 100% window would
measure the average of the entire positive or negative pulse. In measure
all edges mode and EZJIT, these measurements can be trended,
histogrammed, etc.
<source>
<N>
{CHANnel<N> | FUNCtion<F> | WMEMory<N>}
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
<pulse width
percent>
Example
FUNCtion<F> is an integer, 1- 16.
pulse width percent to use in average (in percent, 0- 100)
This example measures the average of the data of a positive pulse within
the pulse window (width set to 50%)
myScope.WriteString ":MEASURE:PTOP CHAN1, 50"
Query
:MEASure:PTOP? <source>, <pulse width percentage>
The :MEASure:PTOP? query returns the average of the data of a positive
pulse within the specified pulse window.
618
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
Related documents
Agilent Technologies Ininiium 90000 Technical data
Agilent Technologies 86100A Technical data
Agilent Technologies 2000 X-Series Technical data
(Agilent) 83495A Optical/electrical clock recovery
VDO MS 2000X - Technical data
NCPlot™ v2.21 User Manual
NCPlot™ v2.26 User Manual
Agilent Technologies InfiniiVision 3000 DSO-X 3054A Technical data
Ditec CROSS18-19 Automazioni per cancelli scorrevoli
Here - FTP Directory Listing