30 Serial Data Equalization Commands
:SPRocessing:FFEQualizer:TAP:TDMode
Command
:SPRocessing:FFEQualizer:TAP:TDMode {TBITrate | CUSTom}
The :FFEQualizer:TAP:TDMode command sets Tap Delay field to either
Track Data Rate or Custom. If you are using one tap per bit, use the
TBITrate selection. If you are using multiple taps per bit, use CUSTom and
then use the FFEQualizer:TAP:TDELay command to set the value. To
understand more about this parameter, consult the N5461A Infiniium
Serial Data Equalization User's Guide.
Example
This example sets the FFE Tap Delay mode to TBITrate.
myScope.WriteString ":SPRocessing:FFEQualizer:TAP:TDMode TBITrate"
Query
:SPRocessing:FFEQualizer:TAP:TDMode?
The :SPRocessing:FFEQualizer:TAP:TDMode? query returns the current Tap
Delay mode.