Download Enable-IT 868 User`s manual
Transcript
December 2003 Low and High Calibration Use the following steps to calibrate the inputs: Calibrating the Low Input 1. Connect the low temperature reference (or current source) to the temperature input (see Figure 8-3 on the previous page). 2. If you have not already, enter the Calibration Menu by pressing the [CAL] key. 3. Press [F2] to select “A/IN.” 4. The prompt will ask you to select Input 1 or Input 2. Press [F1] to select Input 1 or [F2] to select Input 2. 5. Depending on your calibration reference, press [F1] to select “LOW” or [F3] to select “4mA.” 6. If you selected “LOW,” enter the low temperature reference value (in either °F or °C depending on the units selected), then press [ENT]. 7. Press [F1] to select STORE or [F2] to select ABORT. The prompt returns you to the start of the A/IN menu. Calibrating the High Input 1. Connect the high temperature reference (or current source) to the temperature input for supply (see Figure 8-3). 2. Depending on your calibration reference, press [F2] to select “HIGH” or [F4] to select “20mA.” 3. If you selected “HIGH,” enter the high temperature reference value (in either °F or °C depending on the units selected), then press [ENT]. 4. Press [F1] to select STORE or [F2] to select ABORT. The prompt returns you to the start of the A/IN menu. Note: If you cannot adjust the low and high reading, repeat the procedures. If you still cannot successfully calibrate, contact the factory. 8-6 Calibrating Analog Outputs and Inputs