Download California Instruments 2253iX Instruction manual

Transcript
Programming Manual
4.11
Compact i / iX Series
Power Measurement Subsystem
This subsystem programs the power measurement capability of the power source. Two
measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition
of new measurement data before returning a reading. FETCh returns a reading computed from
previously acquired data.
Individual outputs of a three-phase source are specified by the setting of INSTrument:NSELect.
Subsystem Syntax
MEASure | FETCh
[:SCALar]
:POWer
[:DC]?
:AC
[:REAL]?
:APParent?
:REACtive?
:PFACtor?
:TOTal?
Returns the dc component of power
Returns real power
Returns VA
Returns VAR
Returns power factor
Returns real 3-phase total power
MEASure:POWer?
FETCh:POWer?
Phase Selectable
These queries return the dc component of the power being sourced at the output terminals in
watts.
Query Syntax
Parameters
Examples
Returned Parameters
Related Commands
MEASure[:SCALar]:POWer[:DC]?
FETCh[:SCALar]:POWer[:DC]?
None
MEAS:POW? FETC:POW?
<NR3>
INST:NSEL
MEASure:POWer:AC?
FETCh:POWer:AC?
Phase Selectable
These queries return the in-phase component of power being sourced at the output terminals in
watts.
Query Syntax
Parameters
Examples
Returned Parameters
Related Commands
MEASure[:SCALar]:POWer:AC[:REAL]?
FETCh[:SCALar]:POWer:AC[:REAL]?
None
MEAS:POW:AC?
FETC:POW:AC?
<NR3>
INST:NSEL
MEASure:POWer:AC:APParent?
FETCh:POWer:AC:APParent?
Phase Selectable
These queries return the apparent power being sourced at the output terminals in volt-amperes.
AMETEK Programmable Power Division
55