Download Agilent Technologies E4438C Installation guide
Transcript
Adjustments External Input Peak Detector Calibration External Input Peak Detector Calibration Description This calibration is used to calibrate the positive trip level of the EXT1 and EXT2 peak detectors located on the A17 Synthesizer (Option 501, 502, 503, 504). The calibration generates DAC values for the modulation comparator DACs which provide a voltage to the window comparator operational amplifiers. When the peak detector level DACs are set correctly, the overmodulation indicator will trigger whenever the voltage to the EXT1/EXT2 input is greater than 1.03 V. The resulting DAC values are saved as a calibration constant in the signal generator’s firmware. Required Test Equipment • Agilent 3458A Option 002 digital multimeter • Agilent 33120A function generator Procedure Figure 4-18 External Input Peak Detector Calibration Setup 1. Connect the equipment as shown above. 2. Preset all of the equipment. 3. Follow the instructions as they appear on the controller’s display. 4- 25