Download Agilent Technologies 34450A 5 1/2 Digit Multimeter Technical data

Transcript
1
Performance Tests and Calibration
Performance Verification Tests
Table 1-4 DC current gain verification test
Input
Function[1]
Range
Quick Check
Error from Nominal 1 year
1A
1A
Q
±1.15 mA
10 A
10 A
±25.7 mA
[1] Select Slow Mode 5½ digit measurement resolution
Q= Quick performance verification test points
Ohms gain verification test
Configuration: 4- W Ohms (CONFigure:FRESistance) or 2- W Ohms
(CONFigure:RESistance)
1 Connect the calibrator to the front panel HI and LO input terminals
as shown in the figure below:
2- W resistance with compensation (For range 100 Ω ~ 100 kΩ )
2- W resistance without compensation (For range 1 MΩ ~ 100 MΩ )
14