1 Performance Tests and Calibration Performance Verification Tests Table 1-7 AC volts verification test Vrms Input Frequency Function[1] Range Quick Check Error from Nominal 1 Year Q ±3 mV 1V 1 kHz 1V 1V 10 kHz 1V ±3 mV 1V 30 kHz 1V ±18 mV 1V 100 kHz 1V ±33 mV 10 V 20 Hz 10 V ±0.11 V 10 V 45 Hz 10 V ±30 mV 10 V 1 kHz 10 V ±30 mV 10 V 10 kHz 10 V ±30 mV 10 V 30 kHz 10 V 10 V 100 kHz 10 V ±0.33 V 100 V 45 Hz 100 V ±0.3 V 100 V 1 kHz 100 V ±0.3 V 100 V 10 kHz 100 V 100 V 30 kHz 100 V ±1.8 V 100 V 100 kHz 100 V ±3.3 V 750 V 45 Hz 750 V ±2.25 V 750 V 1 kHz 750 V ±2.25 V 750 V 10 kHz 750 V ±2.25 V Q Q ±0.18 V ±0.3 V [1] Select Slow Mode 5½ digit measurement resolution Q= Quick performance verification test points 18 34450A Service Guide