Download Agilent Technologies 34450A 5 1/2 Digit Multimeter Technical data
Transcript
Performance Tests and Calibration Performance Verification Tests 1 Performance Verification Tests The performance verification tests are recommended as acceptance tests when you first receive the instrument. The acceptance test results should be compared against the one year test limits. After acceptance, you should repeat the performance verification tests at every calibration interval. If the instrument fails performance verification, adjustment or repair is required. Adjustment is recommended at every calibration interval. If adjustment is not made, you must establish a 'guard band', using no more than 80% of the specifications, as the verification limits. Ensure that you have read the “Test Considerations” on page 5 before running the performance verification tests. NOTE Zero offset verification This test is used to check the zero offset performance of the instrument. Verification checks are only performed for those functions and ranges with unique offset calibration constants. Measurements are checked for each function and range as described in the procedure on the next page. 34450A Service Guide 9