6000_series_service_guide.book Page 54 Friday, March 23, 2007 2:43 PM 2 Testing Performance Table 10 Equipment Required to Verify Bandwidth Equipment Critical Specifications Recommended Model/Part Signal Generator 100 kHz - 1 GHz at 200 mVrms Agilent E4400B/8648A Power Meter/Sensor 1 MHz - 1 GHz ±3% accuracy Agilent E4418B/8482A Power Splitter outputs differ by < 0.15 dB Agilent 11667A Cable Type N (m) 24 inch Agilent 11500B Adapter Type N (m) to BNC (m) Agilent 1250-0082 Feedthrough 50Ω BNC (f) to BNC (m) Agilent 0960-0301 1 Connect the equipment (see Figure 5). a Use the N cable to connect the signal generator to the input of the power splitter input. b Connect the power sensor to one output of the power splitter. c For 300 MHz, 500 MHz, and 1 GHz oscilloscope models use an N- to- BNC adapter to connect the other splitter output to the channel 1 input. d For 100 MHz oscilloscopes, use an N- to- BNC adapter and 50Ω feedthrough termination to connect the other splitter output to the channel 1 input on the oscilloscope. 54 6000 Series Oscilloscopes Service Guide