Download Advanced Test Equipment Rentals
Transcript
Configuring the Instrument Setting Measurement Method The Method setting allows you to choose between a faster measurement with fewer data points or a slower measurement that contains more data points. The more data points in the measurement, the higher the resolution of the measurement, but the longer it takes to acquire the measurement. Figure 2-- 22 shows the location of the measurement Method function. Method drop-down list Figure 2- 22: Location of Method drop-down list To change the method setting, select the desired setting from the Method drop-down list. Table 2-- 14 shows how the Method setting affects the resolution. Table 2- 14: Performance versus Resolution Method Maximum Resolution for all measurements except Insertion Gain/Loss1 Average Resolution for Insertion Gain/Loss1 Fast 256 100 Normal 1024 200 High Resolution 4096 400 1 2- 56 Resolution is the number of data points acquired per measurement. The actual number of data points depends on several interacting factors. The number of data points displayed may be less than the maximum. YBA250 Antenna and Transmission Line Analyzer User Manual