Section 3.8 Troubleshooting 3.8 Troubleshooting Symptom Remedy Symptom Remedy Symptom Remedy Thermo Scientific Single-Use Mixer will not operate. Check power supply. • Verify main electrical plug connection at wall outlet, verify position of main power disconnect and verify position of Emergency Stop switch. • Verify condition of main electrical circuit breaker of facility. If protection breaker is tripped, determine fault condition. The condition may exist where other electrical systems are requiring current loads beyond those allowed by the breaker. S.U.M. system should be placed on its own electrical circuit. • Disconnect main power cord. Inspect electrical circuit breakers and fuses inside the electrical enclosure of the S.U.M. controller. Determine fault condition by visual inspection. If fault condition can not be determined by visual inspection, contact manufacturer. Speed controller does not respond to user inputs Allow speed to stabilize before using keypad interface • Rapidly adjusting speed control in an excessive manner may require several seconds for speed stabilization. • Wait ten seconds and then attempt to adjust speed at keypad interface. There is too much tension in the film near the bearing port of the loaded BPC. Reload the bag (if possible) by carefully following the instructions in this user’s guide. • Provide excess film to that region of the BPC nearest the bearing port when aligning the bag in the tank. • Do not begin liquid fill until the bag is properly seated in the tank. • Verify the black bumper located in the motor mount block is present. Single-Use Mixer 129