Section 2.3
S.U.M. BPC Loading Instructions
Figure 2.32
6.
7.
Once the bag is fully seated, check that the drain line remains fully
extended through the bottom port before proceeding to the next
step. To keep the drain line from contacting the floor, the line can
be guided through one of the bars located on the base of the outer
support container.
Attach the impeller sleeve directly to the reusable bearing port via
the quick connect attachment on the plastic sheath. An audible click
will confirm that the attachment is made. (Figure 2.34). NOTE: For
the 2000 L unit, make the tri-clamp connection (without a gasket)
using a plastic clamp.
Figure 2.34
8.
Powder Addition
with the Thermo
Scientific HyClone
Powdertainer II
Thermo Scientific
Figure 2.33
Figure 2.35
Insert the drive shaft into the motor as instructed in the previous
section.
The Single-Use Mixer can be equipped with a Powdertainer bag support
stand for solids addition (see Section 4.3 for ordering information).
The following instructions highlight the HyClone Powdertainer BPC in
conjunction with the S.U.M. when equipped with the bag support stand
and holding arm. Other methods for solids addition can also utilize the
powder port. Complete protocols are not included here as individual
protocols vary; however, general descriptions of these processes are
Single-Use Mixer
37