Installation
To program outputs connected in series, first program the current limit
of each output to the total desired current limit point. Then program the
voltage of each output so that the sum of both voltages equals the total
desired operating voltage. The simplest way to accomplish this is to
program each output to one half of the total desired operating voltage.
NOTE
The operating mode of each output is determined by the output’s
programmed settings, operating point, and load condition. Because
these conditions may change during series operation, the operating
status indicators on the front panel will reflect these changes. This is
normal. Momentary status changes are also normal.
Effect on Specifications
Specifications for outputs operating in series can be obtained from the
specifications for single outputs. Most specifications are expressed as a
constant or a percentage (or ppm) plus a constant. For series operation,
the percentage portion remains unchanged while constant portions or
any constants are changed as indicated.
Voltage
All series specifications referring to voltage are twice the single output
specification except for programming resolution, which is the same as
for a single output.
Current
All series specifications referring to current are the same as for a single
output except for CC load effect, CC load cross regulation, CC source
effect, and CC short term drift. These are twice the current
programming accuracy (including the percentage portion) at all
operating points.
Load Transient
Recovery Time
Load transient specifications are typically twice the single output.
Additional Load Considerations
Response Time with an External Capacitor
When programming with an external capacitor, voltage response time
may be longer than that for purely resistive loads. Use the following
formula to estimate the additional up-programming response time:
Response Time = (Added Output Capacitor)X(Change in Vout)
(Current Limit Setting)−(Load Current)
Note that programming into an external output capacitor may cause the
DC Power Analyzer to briefly enter constant current or constant power
operating mode, which adds additional time to the estimation.