Specifications Recovery time Sampler temperature recovers to 4 °C within 5 minutes after the door has been held open for one minute in 24 °C (75 °F) ambient environment while in an active cooling cycle. Pull down time Air temperature drops from 24 °C (75 °F) to 4 °C (39 °F) within 20 minutes (typical). Certifications North America: cETLus listed -Conforms to UL 61010-1, Certified to CSA C22.2 No. 61010-1 and UL 471 and CSA C22.2 No. 120. Europe / IEC: CE - EN / IEC 61010-1 and EN / IEC 60335-2-89 (safety), EN / IEC 61326 (EMC) & CISPR 11 (RF emissions) Sample bottle capacity (see Figure 3 on page 11 and Figure 4 on page 12) Single bottle mode (1) 21 L (5.5 gal) polyethylene or (1) 10 L (2.5 gal) polyethylene or (1) 10 L (2.5 gal) glass bottle Multiple bottle mode (2) 10 L (2.5 gal) polyethylene and/or 10 L (2.5 gal) glass bottles (4) 10 L (2.5 gal) polyethylene and/or 10 L (2.5 gal) glass bottles (8) 2.3 L (0.6 gal) polyethylene and/or 1.9 L (0.5 gal) glass bottles (24) 1 L polyethylene and/or 350 mL glass bottles Intake accessories Strainers 316 stainless steel in standard size, high velocity or low profile for shallow depth applications and Teflon®/316 stainless steel in standard size Sample intake tubing 9.5 mm (3/8 in.) I.D. vinyl or Teflon®-lined polyethylene SD900 controller Enclosure High-Impact, injection-molded PC/ABS blend; submersible, watertight, dust-tight, corrosion, and ice resistant; NEMA 4X, 6, IP 67 Power requirements Refrigerated sampler: 15 VDC provided by 8754500 power supply AWRS sampler: 15 VDC provided by integral power supply Overload protection 7 amp DC line fuse for pump Pump Peristaltic high speed, with spring-mounted rollers Pump rollers Nylatron, impact/corrosion resistant Pump enclosure Track is injection-molded polyphenylene sulfide. Cover is polycarbonate, high impact resistant. Pump enclosure rated IP37. Pump tubing 9.5 mm I.D. x 15.9 O.D. mm (3/8 in. x 5/8 in.) silicone 20,000 sample cycles under the following conditions: • 1 L sample volume • 1 rinse Pump tubing life • 6 minute pacing interval • 16 ft of 3/8 in. intake tube • 15 ft of vertical lift • 70 °F sample temperature Tubing replacement time < 1 minute using pre-cut pump tube Maximum vertical lift to draw sample Minimum of 28 ft, using 29 ft of 3/8-in. vinyl intake tube at sea level at 20–25 °C (68–77 °F) Pump flow rate 1.25 gpm (4.8 L/min) at 3 ft (1 m) vertical lift using 3/8-in. intake tube Typical sample volume repeatability ± 5% of 200 mL sample volume using uncalibrated liquid detect with 15 feet vertical lift,16 feet of 3/8-in. vinyl intake tube configured for a single bottle using full bottle shut-off at room temperature and 5000 ft elevation Typical sample volume accuracy ± 10% of 200 mL sample volume using uncalibrated liquid detect with 15 feet vertical lift,16 feet of 3/8-in. vinyl intake tube configured for a single bottle using full bottle shut-off at room temperature and 5000 ft elevation 6