Download User Manual - Liu Research Group
Transcript
Capillary Replacement Figure 4-8. Capillary replacement dimensions. NOTE The drawing in Figure 4-8 is a representation of the capillary configuration and does not depict the capillary when in use. Tolerance on the capillary dimensions is ± .06” (1.5mm). NOTE Use safety glasses when uncoiling and cutting the capillary. 1. Uncoil the capillary carefully and cut to a length of 7.5 inches (190mm) using a capillary precision cutting tool (SGE #0625010) or comparable tool. 2. Mark insertion dimensions on each end using tape or permanent marker. Mark the end that you insert into the ferrule on the inlet flange 2.2 inches (57mm) from the end. Mark the end you insert into the ferrule on the flowby tee 0 .9 inches (22mm) from that end. 3. Loosen the nut holding the ferrule on the inlet flange to release the ferrule (Figure 4-8). Thread the capillary through the ferrule. 4-16 | AMETEK ProLine Mass Spectrometer