4. Insert the ferrule and capillary onto the inlet flange and secure with the nut
just to keep in place.
5. Adjust the capillary to the insertion point you marked by loosening the nut
to allow the capillary to move in and out easily.
6. Once the capillary is in the correct position (at the insertion point you
marked), tighten the nut until the capillary is grasped and then tighten 1/4
turn more. Do not overtighten.
7. Repeat the insertion process for the ferrule on the flowby tee.
8. Once the capillary is secure in both ferrules and at the correct insertion
points, remove any temporary markers or tape.
Maintenance & Troubleshooting | 4-17