Download - Bio-Rad
Transcript
Lavage, Sputum, and Other Biological Fluid Samples Keep all samples on ice until ready for use. 1. If dilution is required, use Bio-Plex sample diluent with 0.5% BSA final. 2. Centrifugation at 10,000 x g for 10 min at 4°C may be needed to clarify the sample. Sample Dilution for Adiponectin and Adipsin Assays Note: Physiological levels of adiponectin and adipsin are typically found at high concentrations; therefore, higher sample dilutions are required to achieve measurable concentrations within the standard curve. 1.For human adiponectin and adipsin assays, dilute serum or plasma 1:400 in serum-based diluent with two 1:20 serial dilutions nFirst dilution: 10 µl sample + 190 µl serum-based diluent n Second dilution: 10 µl from the first dilution + 190 µl serumbased diluent 2. For mouse and NHP adiponectin assays, dilute serum or plasma 1:1,600 in serum-based diluent with two 1:40 serial dilutions nFirst dilution: 10 μl sample + 390 μl serum-based diluent n Second dilution: 10 μl from the first dilution + 390 μl serumbased diluent 19