Rational Imaging Installation Manual
FACILITY:
M.
ü
ü
ü
ü
ü
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
q
05.08.2001
DATE:
INSTALLED BY:
FILM (LASERLINK) PRINTER
SparcStation compatible with Card:
Kodak (3M) engineer notified:
Kodak (3M) digital port purchased and installed:
Kodak (3M) digital port configured to Cemax specs:
Cables extending from Sun to Kodak (3M):
Shutdown system
Install LaserLink card in SCSI slot 0 (bottom nearest processors)
Install cables to LaserLink card and to Kodak (3M)
Reboot with –r to have system recognize new hardware
Install ISTfilm package
Change directory to /opt/ISTri/film
Check the README file for any system patches that need to be applied.
Install drivers “/opt/ISTri/film/INSTALL
Run /opt/ISTri/film/addcam in that directory to create camcap file
Edit camcap and change DefFormat=”1 on 1”
Edit camcap and change MaxHSize and MaxVSize = 6000 (under 1 on 1).
Link camcap (“ln -s /opt/ISTri/film/camcap /opt/ISTri/bin/camcap”)
Link filmer (“ln -s /opt/ISTri/film/filmer /opt/ISTri/bin/filmer”)
Test printer with filmer “filmer –c camcap kodak.fjob1”
License rifilm for that workstation (add_license < licfile)
Start rifilm in the background using load_riconnect (it will watch /img/film directory)
Check world r+w permission on /img/film
Create a new printer (film printer) in risetup on server
Use RIview to send some images to the Cemax Laserlink (/net/cemaxhost/img/film).
Make sure the images in /img/film are printed and deleted.
If problems persist, try setting the Cemax card with Odd parity.