M mode
4B mode
Figure 3-6. Five display modes sketch
Adjusting keys
Adjusting keys are used to adjust focus position, dynamic range, scanning angle, zoom, cine loop See” Chapter 4
Operation Procedures”)
Focus number selection
Press this key to lighten up the “ focus number” on the screen right top, press again to change the focus numbers.
There are two focus at most. Press this key to switch between one focus to two.
Focus space selection
When set the focus number as two, their distance can be adjusted. There are 5 levels focus space 2, 3, 4, 5, 6 and
press this key repeatedly to switch among them circularly.
Scan angle selection key
Press this key to lighten up the “Angle” on the up right screen , press
or
to change the scanning
angles. The angel range is 50%-100%.
Tips: When linear probe or rectal linear probe is used, this function can not be used.
the angle can improve the image’s frame rate and increase the cine loop frames
Reduce
Dynamic range adjusting key
Press this key to lighten up “DR” on the screen top right corner. Press
or
key to adjust the
dynamic range between 100dB-131dB. 118 dB is recommended.
Depth Adjusting key
Press
or
keys to adjust scan depth between 40mm-240mm. The current depth displays at bootom
screen (see figure 3-1). The depth adjusting range of each probe is given below:
Probe mode
Depth
range(mm)
60R/3.5MHz
Convex probe
70~240
7.5 MHz
Endo-rectal probe
7.5MHz
HF linear probe
40~90
20R/5.0MHz
Micro
convex
probe
80~150
Table 3-1. Probes and their depth adjusting range
19
13R/6.5MHz
Endo-cavity probe
50~120