Download Chapter One Summary
Transcript
TABLE: Summary of the acoustic quantities 60R/3.5MHz convex probe (B+M)-Mode Index MI Mode TIS Scanning Maximum index value 0.50 Acoustic working frequency (MHz) 2.91 TIS TIS Nonscanning Nonscanning Aaprt≤ 1cm2 Aaprt> 1cm2 TIB Scanning 0.061 2.91 2.91 0.29 Bounded output Power (mw) Attenuated output Power (mw) Nonscanning 0.061 3.34 Output power (mw) TIB 4.42 3.34 0.29 4.42 0.20 0.20 Spatial-peak temporal-average intensity(mw/cm2) 2.39 2.39 Attenuated spatial-peak temporal-average intensity(mw/cm2) 0.45 0.45 Pulse-pressure squared integral(mj/cm2) 0.046 0.021 Attenuated Pulse-pressure squared integral(mj/cm2) 0.031 0.0049 Peak-rarefactional acoustic pressure (MPa) 1.03 Attenuated peak-rarefactional acoustic pressure (MPa) 0.85 -1 2dB output beam area (cm2) 1.33 Equivalent aperture diameter (cm) 1.30 Depth for TIS (cm) 1.96 Depth for TIB (cm) Depth at max. attenuated pulse-intensity integral (cm) 1.70 5.55 Supplementary information: 52 TIC