User Manual
•
•
•
•
•
•
<freq> sets the output frequency in units of hertz
<ampl> sets the output amplitude in units of volts
<offs> sets the output offset in units of volts
<delay> sets the delay to the pulse rise transition in units of %
<rise> sets the pulse rise time in units of %
<fall> sets the pulse fall time in units of %
Parameter type
Numeric
Parameter range
<freq>
10e3 to 50e6
<ampl>
10e-3 to 10
<offs>
-4.5 to +4.5
<delay>
0 to 99.9
<rise>
0 to 99.9
<fall>
0 to 99.9
APPLy:RAMP?
Response
The 397 will return <freq>, <ampl>,<offs>,<delay>,<high>,<fall>. Default values are 1e6,5,0,10,10,10
APPLy:SINC{<freq>,<ampl>,<offs>,<N_cycles>}
Purpose
This command is a high-level command that programs the 397 to output sinc waveform along with its
associated parameters
•
•
•
•
<freq> sets the output frequency in units of hertz
<ampl> sets the output amplitude in units of volts
<offs> sets the output offset in units of volts
<N_cycles> sets the number of zero crossing
Parameter type
Numeric
Parameter range
<freq>
10e3 to 50e6
<ampl>
10e-3 to 10
<offs>
-4.5 to +4.5
<N_cycle> 4 to 100
APPLy:SINC?
Response
The 397 will return <freq>, <ampl>,<offs>,<N_cycle>. Default values are 1e6,5,0,10