Download Gradient Pump
Transcript
Young Lin Gradient Pump 8) Change of Pressure Display Scale Display Pressure Scale : 1.00 Input range 0.5-2.0 In case pressure displayed is not same with real pressure, input the difference value for calibration. 9) Selection of DA Voltage Output Display Chart out : Pressure Selection Pres., %A, %B, %C, Flow, Visco, Curve, SL-CH1, SL-CH2 Output of pressure, mixing rate, flow rate, viscosity, gradient curve and frequency ratio between channel 1 and channel 2 via analog output terminal. 10) Setup of DA Voltage Output Offset Display Chart Offset : 420 (0-1000) Input range 0 - 9999 Input of offset value for DA voltage output 11) Setup of DA Voltage Output Gain Display Chart Scale: 1.00 Input range 0.01 - 9.99 Input gain value of DA output 12) Change of Zero Point for Pressure Display Pressure Zero : No Input range Yes, No Changes the zero point of current pressure output. As the offset value of a pressure sensor can be largely changed by temperature and pressure, recommended to re-setup the zero offset after drop pressure. 13) Setup of Auto-flushing Display Rinse Operate: Yes Selection Yes, No Select "Yes" if you want to clean the Gradient Pump inside by an auto-flushing pump built-in inside. It flushes every 3 minutes. 1-31