37 PHODS Checklist Insert DISPLAY AND ALARM INFORMATION Normal Operation - Green LED single flash per breath during inhalation - When you set the unit to ON mode, you must wear your cannula or face mask. Even though you are below the 10,000 feet threshold, the Apnea alarm is “ON” and will detect that no one is breathing, causing the alarm to activate with the flashing Green LED. The unit is calibrated for up to 45 BPM (Breaths per Minute) in all active switch positions. FLOW FAULT ALARM Coinciding with your inhalation, the OPC M1 will produce a two second flashing Green LED to indicate that the oxygen flow from the regulator has stopped. APNEA ALARM Green LED (4 flashes) every four seconds to indicate that breathing has not been detected within the past 30 seconds and will continue to flash until breathing is resumed. LOW BATTERY The OPC M1 provides 3 levels of low battery warning: (1) The first warning Green LED will single flash every 2 seconds. (2) The second warning Green LED will double flash every 2 seconds. (3) Within the last approximate 1 hour of battery life, the oxygen flow will stop BATTERY CHECK When OPC M1 is turned on- Green LED will flash rapidly and an audible alarm will sound and cease - verifies the batteries and circuits are working. If the Batteries are low; LED will continue to flash, Alarm will continue to sound & Solenoid valve will continue to cycle- Replace Batteries. Severely depleted batteries will only flash the LED. Figure 49 Pilot’s or crew’s knee board (See Fig. 49). Copy and cutout for use