Download Tabor Electronics Ltd.

Transcript
5061/5062
User Manual
Offset for the 5062 can be adjusted with 6 digits of resolution
(nearly 19 bits) and amplitude is programmed with 4 digits.
Amplitude and offset can be programmed independently as long as
the following relationship between the two values is not exceeded:
Window ≥
Amplitude
+ Offset
2
The first thing you do before you program amplitude and offset
setting is define which of the channels is being programmed. The
active channel is displayed at the upper right corner of the LCD
display.
When the display shows
at the upper right corner, you are
currently programming channel 1 parameters. Keypads “1” and “2”
are used as hot keys for channel selection. While not editing any
parameter, press key “2” to program channel 2 parameters.
When the display shows
at the upper right corner, you can
proceed with channel 2 programming.
The amplitude and offset parameters are duplicated in multiple
screens however, when changed for a specific function shape, the
new value is updated on all screens for all other function shapes.
Refer to Figure 3-8 and modify amplitude and offset using the
procedure as described below. The index numbers in Figure 3-8
correspond to the procedure steps in the following description.
1.
2.
3.
4.
5.
Press the Amplitude soft key button
Press Enter to edit the Amplitude value
Use the numeric keypad to program the new value
Press “m” for mV, or “x1” for volts to select the suffix letter.
Press Enter to lock in the value
Alternately, you can modify the amplitude value with the dial and
arrow keys but then the termination of the process is by pressing
Enter.
Offset is programmed the same way as amplitude except select
Offset from the soft key menus to access the offset parameter.
Note
If you use the dial or arrow keys to modify the amplitude
or offset parameters, the output is updated immediately
as soon as you modify the parameter. The final value will
be locked in as soon as you press Enter. If you choose
to leave the old value, press Cancel to terminate the
process and to discard of any change made to this
parameter
3-14