Transcript
2. Inspect the o-rings on the valve stem. Replace any that show signs or damage or wear. Valve Stem Place O-rings Here 3. Blow air through the valve stem to make sure it is not plugged. 4. Screw the valve stem into the Tension Post and tighten it completely. 5. Determine the safe volume and initial pressure for the test temperature. Refer to the chart on page “Specifications” on page 4. 6. Fill the cell with fluid. Do not overfill the cell. 7. Clean any spilled fluid from the edge of the cell and place the inner cap assembly onto the cell body. Valve Stem (#175-016) Inner Cap Assembly Cell Body (#175-80-010) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 6