Sensor calibration requires that you provide two reference solutions. One solution
will be the ‘zero’ (sometimes called the ‘blank’) solution and will contain none of the
material being sensed. The other solution, the ‘calibration’ solution, will contain a
known quantity of the material being sensed. It may also be that the calibration is a
solid standard. This calibration quantity must be no more than the measurement
range of the sensor on its 1X scale but it can be less. See Chapter 1 for a description
of sensor ranges.
Sensors are calibrated in response to the Calibrate button, after both the Unit of
Measure and the Calibration Value are entered.
Enter the unit of measure that is appropriate for the Cyclops 7 sensor mounted on the
logger.
Enter the calibration value for the solution.
Click on the Calibrate button. The adjacent scroll bar shows how far the program has
progress through the calibration procedure.