Download E1465A/E1466A/E1467A Relay Matrix Switch Modules
Transcript
Setting the Logical Address Switch The logical address switch (LADDR) factory setting is 120. Valid address values are from 1 to 255. The matrix module can be configured as a single instrument or as a switchbox. See Figure 2-2 for switch position information. NOTE The address switch selected value must be a multiple of 8 if the module is the first module in a switchbox used with a VX/bus command module and is being instructed by SCP/ commands. Logical Address = 120 128 64 Logical Address Switch Location 32 8+16+32+64=120 16 8 4 2 OPEN CLOSED 1 CLOSED = Switch Set To 1 (ON) OPEN = Switch Set To 0 (OFF) Figure 2-2. Setting the Module Logical Address Setting the Interrupt Level The matrix module generates an interrupt after a channel has been closed. These interrupts are sent to, and acknowledgements are received from, the command module (such as an E1406A) via the VXIbus backplane interrupt lines. For applications where the matrix module is installed in a C-Size mainframe and is a servant of the command module, the interrupt line jumper does not have to be moved. See Figure 2-3 to change the interrupt line. You can select seven different interrupt line levels. Line X disables the interrupt and should not be used. The module's factory setting is line 1. To change the setting, remove the four-pin jumper (part number 1258-0247) from the old line location and reinstall the jumper in the new line location. If you are setting the interrupt line to something other than 1, see the E1406A Command Module User's Manual for additional information. If the four-pin jumper is not used, the two jumper locations must have the same interrupt line selected. Chapter 2 Configuring the Matrix Modules 21