Download B&K 9130 User Manual
Transcript
Command syntax: CALibration:SECure:CODE <parameter> Parameter: the password (1 through 4digits) Example: CAL:SEC:CODE "1234" CALibration:STRing This command sets the calibration information. Command syntax: CALibration:STRing <parameter> Parameter: The calibration information about time and so on. Example: CAL:STR "2005-1-9 20:12" CALibration:STRing? This command reads the calibration information. Command syntax: CALibration:STRing? Return parameter: the calibration information saved in the EEPROM. Example: CAL:STR? CALibration:SAVe This command saves calibration coefficients into EEPROM Command syntax: CALibration:SAVe Example: CAL:SAV Note: The calibration parameter will take effect only after it is saved. CALibration:INITital This command renews the calibration coefficients and set the default values. Command syntax: CALibration:INITital Example: CAL:INIT Note: if the calibration fails, you can use this command to renew. 42