Transcript
Issue: Beta - Issue 1 Specification Weight iPCM12A – standard open unuit approx. 300 kg (660 lb) iPCM12B – standard closed unit, no doors approx. 500 kg (1100 lb) iPCM12B – with 2 doors iPCM12C – includes gamma option and ful lead shielding Shielding (for Gamma Monitoring Version) - Option approx. 1700 kg (3750 lb) Booth front and back shield consist of 20 mm of lead. Environmental Temperature and Humidity Operational Temperature Range +0°C to +45°C (32°F to 113°F Storage Temperature Range -10°C to +50°C(14°F to 122°F Humidity Range up to 95% RH non condensing NOTE: USE BELOW +0°C (32°F) IS NOT RECOMMENDED – THE LCD DISPLAY WILL BECOME "SLUGGISH" BELOW THIS TEMPERATURE. DO NOT STORE ABOVE +60°C (122°F) Magnetic Shielding There is no magnetic shielding. Large external magnetic fields may reduce the measured value. IP rating IP50 Environmental restrictions • Not for use in flammable or explosive atmospheres • For installation in "drip-free" locations only • Do not expose to excessive dust pollution levels Thermo Fisher Scientific iPCM User Manual 3-11
Related documents
BENDIX SD-03-1000 User's Manual
GARTEN KULTIVATOR - Lidl Service Website
LASER-WASSERWAAGE
ILLUSTRATED PARTS MANUAL Two Stage Snow
EPLNT-Piping - User Manual V2014.0 - EPLANT
Kompernass KH 3236 Operating Instructions
Access newsletter, Volume 2, issue #1
for Europe: 2.24 MB
User`s Manual
Silvercrest KH 6522 Operating instructions
old version - Electronic Systems and Networking Group
Les robots s`intègrent dans l`usine du futur