30 User Manual | DELTAGuard | Trouble-Shooting 6.1 Error diagnostics fume cupboard function display The following chapters contain notes for error diagnostics and error repairs. To repair some errors special tools are required and/or special knowledge is needed. In case of such errors please get in contact with the Wesemann service department or with the Infralab GmbH. Error diagnostics fume cupboard function display Error Light emitting diodes of the extract air function display do not shine Light emitting diodes oft he extract air function display are flashing possible cause n No voltage n Check voltage supply, if need be, restore voltage supply n Defect connection cable n Replace connection cable n Setting of wrong display type n Defect connection cable n Contact infralab n Replace connection cable n Incorrect calibration n Incorrect orifice factor Actual value through service n Incorrect connection of measuring module not identical to the externally measured value Remedy hose n Contact infralab n Contact infralab n + connection on the sensor remains free - connection on the sensor to be connected LED (LOW) always red n Defect connection cable n Contact infralab n Extract air volume too low n No voltage on the binder for n Check ventilator on site n Inform Infralab motor n Minimum set values ( wrong n Inform Infralab setting) n Inform facility manager and/or LED (LOW) flashing red n Power failure LED (OK) always green n Minimun set values (wrong setting) n Inform Infralab LED (OK) flashing green n Maximum set values (wrong setting) n Inform Infralab Infralab