Download Instructions for use
Transcript
16.3 Oxygen output, inhalation Inlet pressure in operation 2.7 bar Max. O2 output 3.0 bar 3.5 bar 6.0 bar 11 l/min 12 l/min 15 l/min 15 l/min Tolerance ±10 % 16.4 Aspiration performance Maximum performance Maximum performance for the Combi MODULE and the Suction MODULE depend on the inlet pressure of the oxygen supply in operation. The measurements relate to an aspirate bottle with a capacity of 2 liters. Inlet pressure Max. aspiration p2 Max. vacuum performance O2 consumption 2.7 bar – 0.40 bar 8.4 l/min 9.3 l/min 3.0 bar – 0.44 bar 8.6 l/min 10.1 l/min 3.5 bar – 0.53 bar 8.8 l/min 11.4 l/min 4.0 bar – 0.58 bar 9.0 l/min 12.2 l/min 4.5 bar – 0.60 bar 9.0 l/min 14.0 l/min 5.0 bar – 0.60 bar 9.0 l/min 15.0 l/min 5.5 bar – 0.60 bar 8.7 l/min 17.0 l/min 6.0 bar – 0.60 bar 8.5 l/min 19.0 l/min Tolerance ±10 %. If the bacteria filter is used, aspiration performance is reduced as a function of humidity by approx. 1 to 3 l/min. 62 EN Technical data