TML87 Instruction Manual Operating Instructions 6.9.4.5 Current Loop Output Adjustment A current loop option is available and can be installed as a retrofit for each of the analog outputs of the analyzer (Section 5.2). This option converts the DC voltage analog output to a current signal with 0-20 mA output current. The outputs can be scaled to any set of limits within that 0-20 mA range. However, most current loop applications call for either 2-20 mA or 4-20 mA range. All current loop outputs have a +5% over-range. Ranges with the lower limit set to more than 1 mA (e.g., 2-20 or 4-20 mA) also have a -5% under-range. To switch an analog output from voltage to current loop after installing the current output printed circuit assembly, follow the instructions in Section 6.9.4.1 and select CURR from the list of options on the “Output Range” menu. Adjusting the signal zero and span values of the current loop output is done by raising or lowering the voltage of the respective analog output. This proportionally raises or lowers the current produced by the current loop option. Similar to the voltage calibration, the software allows this current adjustment to be made in 100, 10 or 1 count increments. Since the exact current increment per voltage count varies from output to output and from instrument to instrument, you will need to measure the change in the current with a current meter placed in series with the output circuit (Figure 6-6). See Table 3-2 for pin assignments of the Analog Out connector on the rear panel. mA IN OUT V OUT + I IN + V OUT - I IN - Recording Device Analyzer Figure 6-6: Setup for Calibrating Current Outputs NOTE Do not exceed 60 V between current loop outputs and instrument ground. 05496 Rev A6 79