Chapter 9 – Troubleshooting
ERRORS
Message
Processor Fail, Act temp sensor error
Remarks
An error is reported when any of the following
conditions are detected:
• The activator temperature read less than 0.5 Deg.
C (Hardware fault)
• The temperature was greater than 43 Deg. C
(Overlimit)
• The temperature exceeded the setpoint by more
than 3 degrees (FW control loop failed)
Processor Fail, Stab temp sensor error
The stabilizer temperature read less than 0.5
Deg. C (Hardware fault)
Processor Fail, Dryer temp sensor error
The dryer temperature read less than 0.5 Deg. C
(Hardware fault)
Processor Fail, Eco Hardware Interface error
The low level driver failed to initialize resulting
error codes range from 2320 through 2324
Processor Error, Interlock
The processor interlock has been detected resulting error code 1901
Processor Error, Activator Level Low
The activator tank AND the activator replenish
tank are both low resulting error code 2001
Processor Error, Stabilizer Level Low
The stabilizer tank AND the stabilizer replenish
tank are both low resulting error code 2003
Processor Busy, Waste container full
The waste tank is full and must be emptied
Processor Error, Water Level Low
The water supply tank is empty