Download System Manual
Transcript
44 IVIS ® Lumina XR Imaging System Manual 10.2 Specifications Electrical Power & Fuses Voltage Available 100VAC, 50/60 Hz - 240VAC, 50/60Hz Power Consumption 200 W NOTE For more on additional electrical power requirements of the IVIS Lumina XR, see Chapter 6, page 19. Environmental - Temperature & Humidity Temperature 15-25° C Humidity 0% to 80% Non-Condensing Type of Use Indoor Lamp and Fuse Lamp 150 W Quartz Tungsten Halogen 21 V EKE lamp life: 400 hrs (“High” setting), 6,000 hrs (“Low” setting) Color temperature 3250 K° Fuse 3A(10) 250VAC Ventilation Requirements Provide sufficient space (minimum 6 inches) behind the fan of the light source module so that airflow is unobstructed. Provide a similar minimum distance behind the IVIS Lumina XR to enable fan cooling as well as adequate room for fiber optic cable routing. Chemicals Required for Operation No chemicals are required for the operation of the IVIS Lumina XR or the fluorescence equipment. Other user supplied chemicals or materials may be required for your specific biological testing procedures. Weight & Dimensions of the Fluorescence Light Source Module Weight 4.94 lbs (2.24 kg) Depth 8.61 inches (21.9 cm) Width 7.27 inches (18.5 cm) Height 4.6 inches (17.7 cm)