Configuration / Parameter setting via TRWinProg
6.3.4 Output value in case of an error
Determines, which data value is to be transmitted in the case of an error. The data
value is output, if the laser can output no more measurement. This is given e.g., if a
beam interruption is present.
Selection
Description
all 0
The position is set to "0"
all 1
All 24 bits are set to '1' (0xFFFFFF or -1)
last valid value
Output of the last valid position
Default
X
6.3.5 Warning bit Temperature from
Determines, starting from which temperature the warning bit or switching output will be
set.
Selection
Description
47
Message, if device temperature ≥ 47 °C
48
Message, if device temperature ≥ 48 °C
49
Message, if device temperature ≥ 49 °C
50
Message, if device temperature ≥ 50 °C
Default
X
6.3.6 Warning bit Intensity under
Determines, starting from which intensity value of the laser beam the warning bit or
switching output will be set.
Input
1 – 100 [%]
Description
Message, if intensity value < Input value
12 %
© TR-Electronic GmbH 2007, All Rights Reserved