Download 3710 Portable Samplers
Transcript
3710 Portable Samplers Section 5 Routine Maintenance 5.4.3 Installing a New Pump Tube The steps below outline the procedure needed to install a new pump tube. WARNING Pump may actuate without warning. To avoid injury, sampler must be off when pump housing cover is removed for inspection or tubing replacement. 1. The pump tube is marked with two black bands which locate the tubing correctly in the detector and pump. To prolong the tubing life, insure efficient pump operation, and accurate samples, place the tubing in the pump correctly. Facing the liquid detector, position the inner edge of the end band against the upper left inlet of the liquid detector. Place the inner band at the lower outlet of the liquid detector. Figure 5-3 shows the correct placement of the tubing. 2. Slip the pump tubing under the rollers so that the pump tubing does not interfere with the installation of the outer pump lid. 3. Replace the outer pump lid and tighten the four thumbscrews. 4. Re-check the position of the marker bands and adjust the tubing if the position of the bands indicate the tubing has slipped. 5. Replace the outer case of the liquid detector. For proper operation of the liquid detector, tighten the thumbscrews securely. 6. Feed the free end of the tube down through the pump tube port in the center section. 7. Turn the center section over again, and feed the end of the tube into the pump tube guide in the float cage as shown in Figure 5-1. The discharge end of the tube should be flush to 1 /16-inch below the end of the guide. 8. Re-install the suction line, as described in Section 5.5. 9. Inspect the length of exposed tube under the center section. There should be no excessive slack in this tube; it should continuously slope downward from the bottom of the center section to the point where it enters the float cage. This prevents low spots in the tube which might not completely drain. If there is excessive slack, adjust the tube under the center section by pulling it from the pump-tube port. 10. Reset the Pump Tube count to zero. Refer to the discussion of Section 3.9.17, on page 3-36. 5-7