Download 21219977A
Transcript
TÜV Rheinland Energie und Umwelt GmbH Air Pollution Control Report on performance testing of the Air Pollution Monitor 2 (APM-2) measuring system manufactured by Comde-Derenda GmbH for the components suspended particulate matter PM10 and PM2.5, Report no.: 936/21219977/A The results of the check of the five criteria given in chapter 6.1 lence check (modules 5.4.9 – 5.4.11) are as follows: Page 167 of 300 Methodology of the equiva- Criterion 1: More than 20 % of the data are greater than 17 µg/m³. Criterion 2: The uncertainty between the candidates is less than 2.5 µg/m³. Criterion 3: The uncertainty between the reference devices is less than 2.0 µg/m³. Criterion 4: All expanded uncertainties are below 25 %. Criterion 5: The slopes for evaluation of the complete dataset are significantly greater than the permissible values for both devices. Other: For both candidates together, the slope is 0.919 and the intercept is 0.327 at an expanded overall uncertainty of 17.68 % for the full dataset. 3195669_2014_936_21219977A_EN.doc