TÜV Rheinland Energie und Umwelt GmbH
Air Pollution Control
Report on performance testing of the Air Pollution Monitor 2 (APM-2) measuring system manufactured by Comde-Derenda GmbH for the components suspended particulate matter PM10 and PM2.5, Report no.: 936/21219977/A
Page 197 of 300
According to the approach of chapter „8.2 Suitability test“ of EN/TS 16450 [9], the corrected
data set for Cologne, winter 2014 was finally added as a fifth data set to the original equivalence test during type approval (refer to Table 43 for PM2.5 and Table 44 for PM10) and it was
checked, if the criteria of the equivalence test are still fulfilled.
Table 49:
Results of equivalence test „Type approval + Cologne, winter 2014“,
SN 3 & SN 4, measured component PM2,5, slope corrected with 0.919
Comparison candidate with reference according to
Guide "Demonstration of Equivalence Of Ambient Air Monitoring Methods", January 2010
Candidate
APM-2
SN
SN 3 & SN 4
Limit value
30
Status of measured values
Slope corrected
Allowed uncertainty
25
µg/m³
%
All comparisons
Uncertainty between Reference
Uncertainty between Candidates
Number of data pairs
Slope b
Uncertainty of b
Ordinate intercept a
Uncertainty of a
Expanded meas. uncertainty W CM
0.54
0.70
SN 3 & SN 4
239
0.981
0.012
0.872
0.209
12.39
µg/m³
µg/m³
not significant
significant
%
All comparisons, ≥18 µg/m³
Uncertainty between Reference
Uncertainty between Candidates
Number of data pairs
Slope b
Uncertainty of b
Ordinate intercept a
Uncertainty of a
Expanded meas. uncertainty W CM
0.64
1.03
SN 3 & SN 4
61
0.953
0.030
1.663
0.932
18.09
µg/m³
µg/m³
%
All comparisons, <18 µg/m³
Uncertainty between Reference
Uncertainty between Candidates
Number of data pairs
Slope b
Uncertainty of b
Ordinate intercept a
Uncertainty of a
Expanded meas. uncertainty W CM
0.51
0.52
SN 3 & SN 4
178
1.069
0.029
-0.010
0.308
16.79
µg/m³
µg/m³
%
3195669_2014_936_21219977A_EN.doc