DATOSPIR TOUCH User’s Manual
32
CALIBRATION ADAPTER
CALIBRATION SYRINGE
1
Power up the DATOSPIR TOUCH and press
main menu.
2
Discharge the syringe once at any range of the following flows
varying between 0-12L/s:
from the
Low flow level: 0,4 – 1,2 L/s
Mid-flow level: 2 - 5 L/s
High flow level: 6 - 12 L/s
To simplify the procedure, the spirometer allows you to carry
out a single maneuver at Mid-flow level: 2 - 5 l / s.
3
Press
to cancel the maneuver or
the next range of flow.
to accept it and go to
511-B00-MU2
•
REV. 1 .06