Download MXI-2 Configuration Reference Manual
Transcript
Chapter 6 VXI-MXI-2/B Configuration and Installation If you change the S1 setting to drive CLK10 out the MXIbus, do not set the W1 jumper setting to receive the MXIbus CLK10. Instead, use the settings shown in Figure 6-4a or 6-4b as appropriate. Do not configure more than one MXIbus device to drive the MXIbus CLK10. Having a second device driving MXIbus CLK10 could damage the device. Caution Figure 6-6 shows the configuration settings for receiving and driving MXIbus CLK10, respectively. MBCLK10 IN S1 OUT MBCLK10 S1 IN a. Receive CLK10 from MXIbus (Default) OUT b. Drive MXIbus CLK10 from VXIbus CLK10 Figure 6-6. Receiving or Driving MXIbus CLK10 Trigger Input Termination You can use switch S5 to terminate the external trigger input SMB with 50 Ω. Figure 6-7a shows the default setting for a nonterminated trigger input SMB. Use the setting of Figure 6-7b to terminate the trigger input SMB. ( TERM = NO) ( TRIGIN = NO) TERM S5 S5 a. Non-Terminated External Trigger Input SMB (Default) TRIGIN b. Terminated External Trigger Input SMB with 50 Ω Figure 6-7. SMB Trigger Input Termination MXIbus Termination The first and last MXIbus devices connected to the MXIbus—whether using a single MXI-2 cable or daisy-chained MXI-2 cables—must terminate the MXIbus. Any MXIbus devices in the middle of a MXIbus daisy chain must not terminate the MXIbus. The VXI-MXI-2/B automatically senses whether it is at either end of the MXIbus cable to terminate the MXIbus. You can manually control MXIbus termination by defeating the automatic circuitry. Use switches 3 and 4 of © National Instruments Corporation 6-9 MXI-2 Configuration Reference Manual