Troubleshooting
Poor Rotation
Lubricate with AR spray.
And then, disconnect handpiece and step on foot pedal.
force?
Does air come out of the end of the tube with enough
No: Defective tube. Contact serviceman.
Yes: If other handpieces work OK, replace the capsule cartridge.
Make sure the capsule is the correct one.
Operation 2011-09-21