Download Low Flow Transmitter EJX115A and EJA115E
Transcript
<7. Operation> 7-13 7.7.13 Local Parameter Setting Lock To disable parameter changes by the Local Parameter Setting there are three different ways. Communication Parameter Ext SW =disable (EXT ZERO ADJ = disable) Communication Parameter Write Protect = On (WRT PROTECT = Yes) Hardware write protection switch on CPU assembly = D (Disable) Locked features • External Zero Adjustment • Local Parameter Setting • Local Parameter Setting • All Communication Parameters * • Local Parameter Setting • All Communication Parameters * *External Zero Adjustment is unlocked. The above parameter setting is carried out by using field communicator or DTM. See the user’s manual IM 01C25T (HART/BRAIN) for the setting procedure. Reviewing local parameter setting by push button on the integral indicator is available at any time even when the Local Parameter Setting is locked. 7.7.14 Others • Difference between BRAIN and HART Tag Number Output Mode (Linear/Sqroot) HART Up to 8 characters can be set “TRNS.FC” shown on the integral indicator BRAIN Up to 16 characters can be set “OUT.MOD” shown on the integral indicator • The degree of adjustment depends on the speed of turning the adjustment screw. Turn the screw slowly for fine tuning and turn the screw fast for quick tuning. IM 01C25K01-01E