Download Technical Manual
Transcript
Proving To simulate actual conditions when testing the Dattus with a 2” flange, attach a 2” pipe/flange combination to the meter’s inlet. Follow Figure 4.2 to properly connect the meter, 25 ft hose, choke valve, dampener, 3 ft hose and transfer prover. Mete r Choke valve 3 ft hose Dampener To Prover 25 ft hose Figure 4.2 Damper Connection Schematic The addition of the “Choke Valve” provides additional pulsation dampening at lower flow rates. For best performance of the dampener, the Choke Valve should be in the “closed” position for calibration below 3000 acfh. Table 4.1 Conditions to use the Damper & Position of Ball Valve Flow Rates >3,000 CFH Transfer Prover Position of Ball Valve Reference Meter 10M Open 3,000 CFH - 500 CFH 2,000 CFH - 300 CFH 10M 2M Closed Closed Connecting the Pressure, Temperature, and Flow Hoses The Dattus meter has three pressure taps on each side of the meter that may be used for testing. 1 Choosing either side of the meter, locate the center pressure tap and the one closest to the inlet of the meter as indicated in Figure 4.3. Dattus fM Meter Technical Reference Guide 4- 2