Download Getting Started - Center for Structural Biology
Transcript
Chapter 7. Acquiring Data The Property button opens a second window, shown in Figure 56, that allows you to set configuration options. The “Status” Concept Every pulse sequence can be divided logically into “periods” of time. The standard two-pulse sequence, for example, can be divided as shown in Figure 57. This sequence has three logical periods, referred to in the diagram as A, B, and C. These periods are used in controlling the decoupler “status” (as well as the “homospoil” status, discussed later in this chapter). For example, say we want the decoupler to be on during period A, on during period B, and off during period C. Using the letter n to signify “no” or an “off-status,” and y to signify “yes” Figure 56. Property Button Window or an “on-status,” we could then (dps Program) describe the desired decoupler status as 'yyn'. Setting dm='yyn' will select this experiment, which in the heteronuclear case might produce a coupled spectrum with NOE, or in the homonuclear case might be used for solvent presaturation experiments. dm='nny' would give us an experiment with the decoupler only on during period C, the acquisition time, which in the heteronuclear case would be a decoupled spectrum without NOE. p1 pw d1 A d2 B at C Figure 57. Logical Periods A, B, C in Standard Two-Pulse Sequence Observe Transmitter and Decoupling Parameters On all systems, transmitter power levels are set through attenuators, which are in turn controlled through a number of parameters. Figure 58 and Figure 59 show schematics for the attenuator configuration for different systems. The parameter tpwr sets the observe transmitter power, which is under computer control on systems with linear amplifiers. tpwr can be given values from 0 to 63, or from –16 to 63, depending on the range of attenuators present in the system. In both cases, 63 is the maximum possible power. The first decoupler mode parameter dm determines first decoupler output: if dm='y', the first decoupler is on, and if dm='n', the first decoupler is turned off. On UNITYINOVA and UNITYplus systems, setting dm to 'a' or 'y' specifies the asynchronous mode. In this mode, the decoupler rf is gated on and modulation is started at random places in the 190 Getting Started: VNMR 6.1B 01-999083-00 A1298