Set up Density, Den. Alarms, & Flow
Density Measurement Setup
●
5202
●
5203 or 5204
●
user’s geometry factor
●
Z-pipe
●
one-piece insertion head (also called a sugar pan or tank probe)
If your gauge head type is not listed, select “user’s geometry factor.” An
additional menu item will be displayed to let you enter a custom
geometry factor. Contact Thermo Fisher for help in determining the
geometry factor for your gauge head type. The default user’s geometry
factor is 0.85.
4. Press the right arrow to scroll through the list of material types: slurry,
solution, single phase, or emulsion. See “Material Type” later in this
chapter for detailed discussion. When the correct material type is
displayed, press the down arrow to accept the selection and move to the
next menu.
Material type is
slurry
CONTINUE CHANGE
Figure 3–4.
5. The wording of the following menu item depends on the material type
selection in the previous item. For a slurry, enter the specific gravity of
the carrier liquid. For a solution, enter the solvent gravity, and so on.
Press the down arrow to move to the next menu item.
Carrier gravity
.9982 g/cc
NEXT
Figure 3–5.