Action Items Hold Action Items 5. The Hold Current Output at FAULT LOW command allows you to hold the current output at the FAULT LOW level (3.6 mA or less). Press the right arrow to execute the command or the down arrow to continue to the next menu item. Hold current outputs at FAULT LOW Exit this menu. NEXT EXECUTE CMD Figure 7–22. 6. The Hold Current Output at FAULT HIGH command allows you to hold the current output at the FAULT HIGH level (20.8 mA or greater). Press the right arrow to execute the command or the down arrow to continue to the next menu item. Hold current outputs FAULT HIGH Exit this menu. NEXT EXECUTE CMD Figure 7–23. 7. Enter the value (in percentage) of the midrange current output hold value. The default value is 50.00%. Press the down arrow to move to the next menu item. Current output hold mode value 50.00% of scale NEXT Figure 7–24 8. Press the right arrow to execute the Hold Current Output command to hold the current output at the midrange value you specified in the previous step. Press the down arrow to continue to the next menu item. Hold current output at 50.00% of scale Exit this menu. NEXT EXECUTE CMD Figure 7–25. Thermo Fisher Scientific DensityPRO Gauge User Guide 7-7