7. TROUBLE SHOOTING
7.1. Instrument failures
Some failures or malfunctions of the instrument are indicated by the applicative software as
error message. Follow the software instructions to solve the problems and take corrective
actions.
7.2. Operating error with possible causes and corrective action
no
Error message
Possible Cause and corrective action
1
It is Not selected Hole
When running the Auto Mode neither the
Donor block nor the Recipient block hole
has been selected during the setting
mode. Please refer to section 5.8
“Process Setting” menu of this operating
manual.
2
It is already registered with Hole.
When during the “ Process Setting” mode
the hole of the Recipient Block has been
selected two times, this message will
show. Please select another hole on the
Recipient block to receive the tissue
specimen.
3
Nothing to extract from a DONOR
When using the Setting Mode, the
at this time.
maximum number of punching holes from
Donor block is 40.
This error message will disappear
automatically and no more selection will
be accepted.
4
Selected All Hole
When the available number of holes in the
Recipient block has been filled and no
more tissue specimens from Donor blocks
can be accepted.
5
Do Initialize
To run the instrument, initialization from
Auto mode must be selected. When this
error message appears, please initialize
from the Auto mode menu screen.
62